Intracellular Calcium[1]                       02.09.02

100x Staining Solution

1µl Fluo-3 AM (10mg/ml in DMSO)            Probes F1241

2µl FuraRed AM (10mg/ml in DMSO)           Probes F3020

27µl Pluronic F-127 (10% in water)         Sigma P2443

(i.e. in 1x solution: 3µM Fluo-3, 6µM FuraRed, 0.1% Pluronic)

 

Cell Stain

1ml      Cell suspension

10µl     100x Staining solution

45 min in the incubator

1x wash with PBS (4°C if possible)

 

Calcium Calibration Curve

Calcium calibration Buffer Kit #1          Probes C-3008

 

Free Ca

Ca free Solution

Ca solution

0µM

500.0 µL

--

17nM

450.0 µL

50.0 µL

38nM

444.5 µL

55.5 µL

65nM

437.5 µl

62.5 µL

100nM

428.5 µl

71.5 µL

150nM

416.7 µl

83.3 µL

225nM

400.0 µl

100.0 µL

351nM

375.0 µl

125.0 µL

602nM

333.3 µl

166.7 µL

 

1µL Calcimycin A23187 (10mM in DMSO) each  Fluka 21045

 

Measurement

Flow Cytometer

Fluo-3              FL-1

FuraRed             FL-3

 

Fluorescence-Reader

Excitation:         485/20nm

Emission Fluo-3:    528/20nm

Emission FuraRed:   590/35nm

(not suitable, EMax at 650nm)

 

from Molecular Probes

 

 



[1] E.J. Novak and P.S. Rabinovitch, Improved sensitivity in flow cytometric intracellular ionized calcium measurement using fluo-3/Fura Red fluorescence ratios. Cytometry, 1994. 17: 135-41.